CRITICAL🇵🇱 Wersja polska

CVE-2014-5432

CVSS 9.8v3.0pub. 2019-03-26upd. 2024-11-21

Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 is remotely accessible via Port 22/SSH without authentication. A remote attacker may be able to make unauthorized configuration changes to the WBM, as well as issue commands to access account credentials and shared keys. Baxter asserts that this vulnerability only allows access to features and functionality on the WBM and that the SIGMA Spectrum infusion pump cannot be controlled from the WBM. Baxter has released a new version of the SIGMA Spectrum Infusion System, Version 8, which incorporates hardware and software changes.

CVSS Vector
CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
  • Baxter Sigma Spectrum Infusion System

    HW
    Baxter
    all versions
  • Baxter Sigma Spectrum Infusion System Firmware

    OS
    Baxter
    6.05
  • Baxter Wireless Battery Module

    HW
    Baxter
    16
🔵
CHECK WITH VENDOR
No clear patch data available. Check vendor references.
Tags
Auth Bypass
CWE
References

Related vulnerabilities

CVE-2020-12047CRITICAL9.8PL ✓same product

Baxter Spectrum WBM — zakodowane na stałe dane logowania w usłudze FTP

CVE-2020-12040CRITICAL9.8PL ✓same product

Baxter Sigma Spectrum — nieszyfrowana komunikacja w systemie infuzji

CVE-2020-12041CRITICAL9.4PL ✓same product

Baxter Spectrum WBM – niezabezpieczony interfejs telnet umożliwia dostęp do danych

CVE-2020-12043CRITICAL9.8PL ✓same product

Baxter Spectrum WBM — niewyłączona usługa FTP po konfiguracji sieci bezprzewodowej

CVE-2020-12045CRITICAL9.8PL ✓same product

Baxter Spectrum WBM — usługa Telnet z zakodowanymi na stałe danymi uwierzytelniającymi