Drager Infinity Delta, Infinity Delta, all versions, Delta XL, all versions, Kappa, all version, and Infinity Explorer C700, all versions. A malformed network packet may cause the monitor to reboot. By repeatedly sending the malformed network packet, an attacker may be able to disrupt patient monitoring by causing the monitor to repeatedly reboot until it falls back to default configuration and loses network connectivity.
CVSS:3.0/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:HDraeger Delta Xl
HWDraegerall versionsDraeger Delta Xl Firmware
OSDraegerall versionsDraeger Infinity Delta
HWDraegerall versionsDraeger Infinity Delta Firmware
OSDraegerall versionsDraeger Infinity Explorer C700
HWDraegerall versionsDraeger Infinity Explorer C700 Firmware
OSDraegerall versionsDraeger Kappa
HWDraegerall versionsDraeger Kappa Firmware
OSDraegerall versions