Drager Infinity Delta, Infinity Delta, all versions, Delta XL, all versions, Kappa, all version, and Infinity Explorer C700, all versions. Log files are accessible over an unauthenticated network connection. By accessing the log files, an attacker is able to gain insights about internals of the patient monitor, the location of the monitor, and wired network configuration.
CVSS:3.0/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:NDraeger Delta Xl
HWDraegerall versionsDraeger Delta Xl Firmware
OSDraegerall versionsDraeger Infinity Delta
HWDraegerall versionsDraeger Infinity Delta Firmware
OSDraegerall versionsDraeger Infinity Explorer C700
HWDraegerall versionsDraeger Infinity Explorer C700 Firmware
OSDraegerall versionsDraeger Kappa
HWDraegerall versionsDraeger Kappa Firmware
OSDraegerall versions
Related vulnerabilities
Dräger Infinity Delta, Delta XL, and Kappa patient monitors contain a denial-of-service vulnerability that all...
Dräger Infinity Explorer C700 contains a privilege escalation vulnerability that allows attackers to break out...